132
Chapter 10
Command Reference
:AVER:COUN
:AVER:COUN
Syntax
[:SENSe]:AVERage:COUNt <numeric>
[:SENSe]:AVERage:COUNt?
Description
Sets the averaging count of the measured value for the averaging function.
Unlike setting the averaging count with the front panel keys, using this command to set the
averaging count does not automatically turn on the averaging function. Therefore, if the
averaging function has been set to off, you have to turn it on by using the
command.
Parameters
If the specified parameter is out of the allowable setup range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
MAX or MIN can be used to specify the parameters.
Query response
{numeric}<newline><^END>
Related commands
Equivalent key
sequence
[Average]
(
[Blue]
-
[Meas Time]
)
<numeric>
Description
Averaging count
Range
1 to 256
Initial value
1
Resolution
1