3-58
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Compliance
Figure 3-9
Relation of Compliance and Output
Current Complinace for COMMON Unit
If you set COMMON output mode for the unit, current compliance for the unit is
automatically set as follows and you cannot change the setting.
GNDU
1.6 A
HRSMU
105 mA
MPSMU
105 mA
HPSMU
1 A
Table 3-6
V Compliance Setting Range
Unit Output
Range
V Compliance
Setting Range
HRSMU
10 pA to 10 mA
0 to 100 V
100 mA (0
£
| I |
£
20 mA)
0 to 100 V
100 mA (20 mA < | I |
£
50 mA)
0 to 40 V
100 mA (50 mA < | I |
£
100 mA)
0 to 20 V
MPSMU
1 nA to 10 mA
0 to 100 V
100 mA (0
£
| I |
£
20 mA)
0 to 100 V
100 mA (20 mA < | I |
£
50 mA)
0 to 40 V
100 mA (50 mA < | I |
£
100 mA)
0 to 20 V
Содержание 4156b
Страница 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Страница 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Страница 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Страница 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Страница 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Страница 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Страница 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Страница 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Страница 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...