To Select a Range
You can let the meter automatically select the range using
autoranging
or you can select a fixed range using
manual ranging
.
•
Autoranging is selected at power-on and after a remote interface reset.
•
Autorange thresholds:
Down range at <10% of range
Up range at >120% of range
•
If the input signal is greater than the present range can measure, the
meter will give an
overload
indication (“OVLD”).
•
When making difference or ratio measurements, an overload on either
channel will give an overload indication (“OVLD”).
For voltage measurements, ranging is local to the selected channel. This
means that you can select the ranging method (auto or manual) for each
channel independently. When manually ranging, the selected range is
local to the active channel; the meter remembers the range when you
switch between channels.
Man
annunciator is on when
manual range is enables
Toggles between autoranging
and manual ranging
Selects a higher range
and disables autoranging
Selects a lower range
and disables autoranging
Chapter 2 Front-Panel Operation
To Select a Range
42
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......