RS-232 Interface Commands
SYSTem:LOCal
This command places the meter in the local mode during
RS-232
operation. All keys on the front panel are fully functional.
SYSTem:REMote
This command places the meter in the remote mode during RS-232
operation. All keys on the front panel except the LOCAL key are
disabled. You MUST send this command before any other RS-232
command can be received.
SYSTem:RWLock
This command places the meter in the remote mode during RS-232
operation. All keys on the front panel are disabled, including the LOCAL
key.
C a u t i o n
Sending or receiving data over the RS-232 interface when not configured
for REMOTE operation can cause unpredictable results. Always make
sure the meter is configured for remote operation when using the RS-232
interface.
Chapter 4 Remote Interface Reference
RS-232 Interface Commands
182
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......