Using the
range
and
resolution
Parameters
With the
MEASure?
and
CONFigure
commands, you can select the
measurement function, range, and resolution all in one command. Use
the
range
parameter to specify a fixed range larger than the expected
value of the input signal. You can set the range parameter to AUTO to
set autoranging.
Use the
resolution
parameter to specify the desired resolution for the
measurement. Setting the resolution sets the integration time of the
meter. The resolution parameter can be set to any number but the
resulting integration time (NPLC) is one of the following:
<
resolution>
Parameter
Integration Time
(NPLC)
< 0.0001 x Range
0.02
< 0.00001 x Range
.2
< 0.000003 x Range
1
< 0.0000022 x Range
2
< 0.000001 x Range
10
< 0.0000008 x Range
20
< 0.0000003 x Range
100
< 0.00000022 x Range
200
Specify the resolution in the same units as the measurement function,
not in number of digits
. For example, for volts measurements, specify the
resolution in volts. For resistance, specify the resolution in ohms.
You must specify a range to use the resolution parameter.
4
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
129
Содержание 34420A
Страница 1: ......
Страница 2: ......
Страница 13: ...1 Quick Start ...
Страница 31: ...2 Front Panel Operation ...
Страница 55: ...3 Features and Functions ...
Страница 117: ...4 Remote Interface Reference ...
Страница 200: ...5 Error Messages ...
Страница 213: ...6 Application Programs ...
Страница 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Страница 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Страница 247: ...7 Measurement Tutorial ...
Страница 274: ...8 Specifications ...
Страница 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Страница 294: ......