84
34410A/11A User’s Guide
2
Features and Functions
Error Conditions
When the front panel ERROR annunciator turns on, one or more
command syntax or hardware errors have been detected. A record of up
to 20 errors can be stored in the instrument’s
error queue
.
•
The instrument beeps once each time a command syntax or
hardware
error is generated.
•
Each remote interface I/O session (e.g., GPIB, USB, LAN) has its own
interface–specific error queue.
•
A special
global
error queue holds all power
–
on and hardware
–
related
errors (e.g., over–temperature).
•
Errors are retrieved in first
–
in
–
first
–
out (FIFO) order. The first
error
returned is the first error that was stored. Errors are cleared as
you read them. Once you have read all of the interface–specific errors,
the errors in the global queue are retrieved.
•
Errors are cleared as you read them. When you have read all
errors
from the interface–specific and global error queues, the ERROR
annunciator turns off and the errors are cleared.
•
If more than 20 errors have occurred, the last error stored in the
queue
(the most recent error) is replaced with
–
350,“Error queue overflow”
.
No additional errors are stored until you remove errors from the queue.
If no errors have occurred when you read the error queue, the
instrument responds with
+0,“No error”
.
•
The front panel reports errors from all I/O sessions as well as
the
global error queue.
•
Both the interface–specific and global error queues are cleared by the
*CLS
(Clear Status) command and when power is cycled. The errors
are also cleared when you read the error queue. The error queue is not
cleared by a Factory Reset (
*RST
) command or an Instrument Preset
(
SYSTem:PRESet
) command.
Calibration
N O T E
Only service–qualified personnel should calibrate the multimeter. Improper
use of the front–panel
CALIBRATION
procedure in the
UTILITY
menu could
result in damage to the multimeter.
Refer to the
Agilent 34410A/34411A Service Guide
for calibration
procedures.
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...