34410A/11A User’s Guide
45
Features and Functions
2
Front Panel Measurement Configuration Menus
Each measurement function is configured separately, except for
W
2W
and
W
4W
(which are partially inter–dependent), and for Frequency and Period,
(which share a configuration menu). The configuration settings (e.g. range,
null value, etc.) for each measurement function are retained when
switching between functions.
•
To enter the configuration menu
for any selected measurement function,
press
. The menus are context–sensitive; only the options
applicable to the selected function will appear. Descriptions of the
menu selections are listed below, for each of the six groups of primary
measurement functions with configurable options (none for
diode/continuity checks.)
•
When digit masking is enabled (
see
on page 43) the
DIGIT MASK
submenu will appear before other configuration choices.
•
For a detailed description of the configuration choices, see
Configuring DC Voltage and DC Current Measurements
For dc voltage and current measurements, your configuration menu
choices are:
INTEGRATION
,
RANGE
,
AUTO ZERO
,
NULL
and
NULL VALUE
. For dc
voltage measurements, you can also configure
INPUT Z
.
•
INTEGRATION
: Allows you to set the integration time for the measurement
in two ways; in power–line cycles (select
NPLC
) or in seconds (select
APERTURE
).
•
RANGE
: Allows you to select a fixed range using
manual ranging
(select
MANUAL
), or let the multimeter automatically select the range using the
autoranging
feature (select
AUTO
).
•
INPUT IMPEDANCE
: Allows you to select a dc input impedance of 10 M
W
(select
10M
W
) or >10 G
W
(select
HI
–
Z
) for the 100 mVDC, 1 VDC and 10
VDC ranges only.
•
AUTO ZERO
: Allows you to enable for all readings (select
ON
), enable for
one reading (select
ONCE
), or disable (select
OFF
) the auto zero feature,
which subtracts a subsequent zero reading from each measurement
taken, to prevent offset voltages in the multimeter’s input circuitry from
affecting measurement accuracy.
•
NULL
: Allows you to enable (select
ON
) or disable (select
OFF
) the null
measurement feature, which measures the difference between a stored
null value and the input signal.
•
NULL VALUE
: Allows you to view and edit the null value.
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...