
34410A/11A User’s Guide
25
Quick Start
1
To Make a 2-Wire Resistance Measurement
•
Ranges: 100
W
, 1 k
W
, 10 k
W
, 100 k
W
, 1 M
W
, 10 M
W
, 100 M
W
•
Maximum resolution: 100
mW
(
on 100
W
range
)
•
Configurable parameters:
INTEGRATION
,
RANGE
,
OFFSET COMP
,
AUTO ZERO
,
NULL
,
and
NULL VALUE
To null
–
out the test lead resistance:
1
Connect one end of the test leads at the meter, and short the probe
ends together.
2
Press null.
3
Connect the probe ends to the test circuit, and measure the corrected
resistance value.
To Make a 4-wire Resistance Measurement
•
Ranges: 100
W
, 1 k
W
, 10 k
W
, 100 k
W
, 1 M
W
, 10 M
W
, 100 M
W
•
Maximum resolution: 100
mW
(
on 100
W
range
)
•
Configurable parameters:
INTEGRATION
,
RANGE
,
OFFSET COMP
,
AUTO ZERO
,
NULL
,
and
NULL VALUE
Содержание 34410A
Страница 1: ...Agilent Technologies Agilent 34410A 11A 6 Digit Multimeter User s Guide...
Страница 6: ...6 34410A 11A User s Guide...
Страница 12: ...12 34410A 11A User s Guide...
Страница 34: ...34 34410A 11A User s Guide 1 Quick Start...
Страница 96: ...96 34410A 11A User s Guide 3 Remote Interface Configuration...
Страница 106: ...106 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 120: ...120 34410A 11A User s Guide 4 Measurement Tutorial...
Страница 135: ...34410A 11A User s Guide 135 Specifications 5 Dimensions All dimensions are shown in millimeters...
Страница 140: ...140 34410A 11A User s Guide 5 Specifications...