38
2200 TapeStation
4
Using the 2200 TapeStation System
Operating Procedure
•
Clicking
Reset All
will remove all manual changes and return the entire
tape to the automatically analysed state.
•
A text file, raw gel image (bitmap), Word and Excel report can all be
exported and saved for each run. In addition, the results can be sent
straight to the printer as a non-editable report. Selecting or de-selecting the
gel view will dictate whether it appears in the Word file and/or
straight-to-print report. The data is a visual representation of the
automatically launched .srf file. In Excel, a positive result is denoted by the
intensity value.
Shutdown and Restarting Procedure
Shutdown Procedure
Ensure that the TapeStation System is shut down in the following order:
1
Exit the TapeStation Controller Software.
2
Turn off the TapeStation instrument.
3
Power down the laptop.
Restarting Procedure
Ensure that the ScreenTape System is restarted in the following order:
1
Power up the laptop.
2
Turn on the TapeStation.
3
Start the TapeStation Controller Software.
N O T E
The controller software, TapeStation instrument and laptop should be shut down when not
in use (preferably at the end of every working day).
Содержание 2200 TapeStation
Страница 1: ...Agilent Technologies Agilent 2200 TapeStation User Manual...
Страница 10: ...10 2200 TapeStation 1 Introduction to the 2200 TapeStation System Overview of the System...
Страница 14: ...14 2200 TapeStation 2 Specifications Technical Specifications...
Страница 42: ...42 2200 TapeStation 4 Using the 2200 TapeStation System DNA Sample Preparation...
Страница 48: ...48 2200 TapeStation 5 Maintenance Changing the Needle...
Страница 64: ...64 2200 TapeStation 6 Troubleshooting Troubleshooting the Test Kit Manufacturers TestKitSuite xml File...
Страница 71: ......