2200 TapeStation
33
Using the 2200 TapeStation System
4
Operating Procedure
Start the TapeStation Run
1
Click the start button.
This will produce a
Save As
window.
As a default the file name starts with the date, in reverse order, and a run
counter. When run continuously, the save function auto increments the
counter part of the file name.
2
Type in the name that you wish the analysis to be saved as. Do not include a
full stop ( . ) in file names.
Final Check
ScreenTapeController:
1
Lift the lid of the TapeStation.
2
Ensure that there are fresh tips in the tip holder and that all the samples
have been correctly loaded with lids removed and correspond to the sample
selection on the screen.
3
Close the lid.
N O T E
Lifting the lid of the TapeStation after this time will abort the experimental run.