Replacing Assemblies
6
16900A, 16902A, and 16903A Service Guide
47
To save the license file
Before doing any major repairs to the instrument, it is
recommended that you back up any data and license files if
possible. You will need a backup copy if you encounter any
problems that require reimaging the hard drive.
The license files have the licenses for the inverse assemblers
and tools. If, for some reason, you lose the license files, you
can obtain new license files from Agilent.
To save the license file:
1
Insert a writeable CD into the CD drive.
2
Use Windows Explorer to copy the folder c:\
ProgramFiles\Agilent Technologies\Logic Analyzer\License
and paste it to the CD.
3
In Windows explorer select the CD- R/RW drive then use
the menu command
File>Write these files to CD
To obtain the licensing host ID
In the Logic Analyzer application, go to
Help>Software
Licensing
. Choose the tab ‘How to install a license’. The
Licensing Host ID will be displayed on this page.