30
16900A, 16902A, and 16903A Service Guide
5
Troubleshooting
Figure 4
16900 and 16902A Boot UpTroubleshooting Flowchart
4
Does the
instrument reach
#3 in the boot
sequence?
Yes
TYPICAL BOOT SEQUENCE DISPLAY
----------------------------------------------------
1. Boot text
2. Agilent Technologies text and logo
3. Intel boot agent text
4. Select operating system menu
5. Windows XP boot screen (blue)
6. Checking file system text
7. Welcome to Windows login screen
Failed BIOS
boot.
Does
the instrument
reach #5 in the boot
sequence and is the
power switch
LED steady
on?
Yes
Does the
instrument reach
#7 in the boot
sequence?
Yes
Enter BIOS setup and
reset to default.
Failed OS boot
from hard
drive.
Replace the CPU
motherboard.
Check the hard drive
power and signal
cables.
Reseat RAM DIMMs.
Run the recovery disk.
Replace the hard drive
and run the recovery
disk.
Failed the system file
check. Run the
recovery disk.
1
Is the
power LED
steady
on?
Blank screen/display problem due to the display mode.
Reboot.
At Select OS menu, press F8 to open advanced startup.
Select "enable VGA mode" then Continue.
Yes
No
No
A
B
C
D
Try , then ,
then . Return to
between each.
C
D
E
E
Try , then .
Return to between
each.
A
B
No
No
1
1
Содержание 16900A
Страница 1: ...Agilent Technologies Agilent 16900A 16902A and 16903A Logic Analysis System Service Guide...
Страница 6: ...6 16900A 16902A and 16903A Service Guide...
Страница 14: ...14 16900A 16902A and 16903A Service Guide 1 General Information...
Страница 18: ...18 16900A 16902A and 16903A Service Guide 2 Preparing for Use...
Страница 22: ...22 16900A 16902A and 16903A Service Guide 3 Testing Performance...
Страница 86: ...86 16900A 16902A and 16903A Service Guide 6 Replacing Assemblies...
Страница 102: ...102 16900A 16902A and 16903A Service Guide 7 Replaceable Parts...