Agilent Technologies 16334A Скачать руководство пользователя страница 1

Agilent 16334A Test Fixture

Operation and Service Manual

Second Edition

Agilent Part No.  16334-90000

December 1999

Printed in Japan

Содержание 16334A

Страница 1: ...Agilent 16334A Test Fixture Operation and Service Manual Second Edition Agilent Part No 16334 90000 December 1999 Printed in Japan ...

Страница 2: ...porated January 1990 First Edition part number 16334 90000 December 1999 Second Edition part number 16334 90000 Safety Summary The following general safety precautions must be observed during all phases of operation service and repair of this instrument Failure to comply with these precautions or with specific WARNINGS elsewhere in this manual may impair the protection provided by the equipment In...

Страница 3: ...atures are maintained Dangerous Procedure Warnings Warnings such as the example below precede potentially dangerous procedures throughout this manual Instructions contained in the warnings must be followed WARNING Dangerous voltages capable of causing death are presenting this instrument Use extreme caution when handling testing and adjusting this instrument Safety Symbol General definitions of sa...

Страница 4: ...f one year Fixtures and adapters are warranted for a period of 90 days During the warranty period Agilent Technologies will at its option either repair or replace products that prove to be defective For warranty service or repair this product must be returned to a service facility designated by Agilent Technologies Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologie...

Страница 5: ... theory Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products For any assistance contact your nearest Agilent Technologies Sales and Service Office Addresses are provided at the back of this manual Typeface Conventions Bold Boldface type is used when a term is defined For example icons are symbols Italic Italic type is us...

Страница 6: ...6 ...

Страница 7: ...Contents 7 1 Operation DESCRIPTION 10 Specifications 12 Typical Characteristics 13 ZERO OFFSET ADJUSTMENT 14 OPERATION 15 2 Service Maintenance 18 ...

Страница 8: ...8 Contents ...

Страница 9: ...9 1 Operation This operation note provides complete information on the 16334A Test Fixture ...

Страница 10: ...a tweezer type probe for measurement of leadless components such as chip capacitors The tweezer tips are replaceable The distance between the tweezer tips is continuously adjustable according to the distance between the DUT terminals A compensation block is furnished for compensation to be minimize the effects of residual impedance and the stray admittance The specifications of the 16334A are give...

Страница 11: ...Chapter 1 11 Operation DESCRIPTION Tweezer Tip 16334 09002 1ea Bumishing Tool 8830 0002 Table 1 1 Accessories Description Agilent Part No ...

Страница 12: ...4276A 4277A or 4278A Measurable Components Miniature components with terminals on opposite faces Distance between component terminals 10mm Terminal Configuration 4 terminal pair configuration from BNC connectors to the top of the tweezers excluding the tweezer tips Cable Length Approximately 1m Maximum Voltage 42V peak max AC DC Measurement Frequency Range 5 Hz to 15 MHz Cable Length Approximately...

Страница 13: ...icable Measurement Range Incremental error 1 MHz f Parameter Value Measurement Frequency Parameter reading error Offset value for D 4192A Full range Full range 2 f 10 2 1 0 02 f 10 2 1 4194A below 15 MHz 4274A Full range 4275A 2 f 10 2 1 0 02 f 10 2 1 4276A 4277A 2 0 02 0 0002 4278A 2 1 There are some cases where the actual incremental error exceeds this limit when this fixture is used with an 419...

Страница 14: ...iduals in order to minimize measurement error the instrument s zero offset adjustment procedure should be performed The zero offset adjustment procedure is described in the instrument s manua1 When performing the zero offset adjustment use compensation block Agilent PN 16334 60001 furnished with probe An illustration showing how to use the compensation block is shown in Figure 1 3 Figure 1 3 Zero ...

Страница 15: ...ctly to the UNKNOUN terminals of the instrument 3 Perform the ZERO offset adjustment described in the instrument s operation manual 4 Connect the DUT to the 16334A test fixture WARNING DO NOT APPLY MORE THAN 42V OF DC BIAS VOLTAGE TO THE UNKNOWN TERMINALS AN ELECTRICAL SHOCK HAZARD WILL EXIST DURING OPERATION WHEN THE DC BIAS VOLTAGE IS GREATER THAN 42V DC ...

Страница 16: ...16 Chapter1 Operation OPERATION ...

Страница 17: ...17 2 Service ...

Страница 18: ...ally of cleaning contacts and replacing worn or damaged parts Take special care when cleaning contacts To order parts use the Agilent Technologies part numbers listed in Table 2 12 1 If a faulty part is located in an assembly that cannot be disassembled order the next higher assembly or return the whole device to the nearest Agilent Technologies Sales Service Office for repair or replacement ...

Страница 19: ...Chapter 2 19 Service Maintenance Figure 2 1 Parts Identification ...

Страница 20: ... 11 16334 04001 1 COVER TOP 12 0515 1550 1 SCREW 13 2190 0206 1 WASHER 14 16047 40000 1 STOPPER 15 1400 0719 2 CABLE TIE 16 16334 40003 2 CABLE BUSHING 17 105 5 cm CABLE 1 18 16334 01202 1 ANGLE 19 16334 24002 1 SCREW 20 2190 0584 1 WASHER 21 16334 0l213 1 BRACKET 2 22 16334 01211 1 BRACKET 2 23 0515 1873 2 SCREW 24 16334 00612 1 ARM PLATE marked L 25 0515 1872 2 MACHINE SCREW 26 16334 40001 2 INS...

Страница 21: ...00601 2 SHIELD 31 0890 0273 2x8 5cm TUBE HEAT SHRINKABLE 32 16334 00613 1 ARM PLATE marked H 33 1400 0249 2 CABLE TIE 1thru33 16334 60010 1 TEST FIXTURE 34 16334 00604 1 PLATE 35 16334 40002 1 BASE 36 16334 60001 1 COMPENSATION BLOCK ASSY 21 OLD 16334 01203 1 BRACKET 2 22 OLD 16334 01201 1 BRACKET 2 Table 2 1 Parts Identification Reference Agilent Part No Qty Description Note ...