2D-LC User Guide
313
11
Error Information
Module-Specific Error Messages
Module-Specific Error Messages
For further module-specific errors, please see the manual of the module in
question.
Initialization of Valve Failed
Error ID: 24000
During the initialization process the motor of the valve drive moves to some
special positions depending on the installed valve head. A failure in this process
means either that the movement couldn’t be performed properly or it was not
noticed correctly by the sensor.
Probable cause
Suggested actions
1
Mechanical problems. Friction too high or
blockages on the valve drive’s motor or on
the valve head.
•
Check valve head for correct installation
•
Try to identify the source of trouble by
installing a different valve head if possible.
•
Contact your Agilent Service representative.
2
Defect Sensor on the Valve Drive Motor
•
Check valve head for correct installation
•
Try to identify the source of trouble by
installing a different valve head if possible.
•
Contact your Agilent Service representative.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...