![Agilent Technologies 1290 Infinity II 2D-LC Solution Скачать руководство пользователя страница 234](http://html1.mh-extra.com/html/agilent-technologies/1290-infinity-ii-2d-lc-solution/1290-infinity-ii-2d-lc-solution_user-manual_2867575234.webp)
9
Data Analysis
2D-LC Data Analysis/Data Evaluation for MassHunter
2D-LC User Guide
234
MassHunter Qualitative Analysis Software
The Masshunter data analysis software generally works with
2
D data in the same
way as you are used to. The task to find an identify compounds or setup and run
qualitative analysis methods can also be performed on 2D-LC data. However, to
make it easier to get started with
2
D data, we have listed some different
workflows as examples.
The 2D-LC instrument in this case was equipped with 2 UV detectors (
1
D and
2
D),
2D-LC valve with MHC and a Q-TOF detector in the second dimension. Default
method is loaded.
Workflow
1
D UV Data Extraction – Alternative 1
1
Open Data File 2D-LC file.d, in this case Test_Comp r001.d.
The string of 2D TIC-chromatograms appears.
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...