2D-LC User Guide
189
7
Method Development of Active Solvent Modulation (ASM)
Optimize the Sample Loop Flush
Optimize the Sample Loop Flush
Activate ASM in the software and set Flush sample loop to 3.0 times.
NOTE
Flushing the sample loop 3 times is typically enough and the recommended
default. Less time may be sufficient and can be verified during optimization. The
user interface displays how long this will take.