2D-LC User Guide
177
6
Method Parameters
Additional Information
Additional Information
Multi-Inject
To sample a broad 1D-region that does not fit into currently installed sampling
loops (e.g. 40 µL volume) HiRes is the method of choice. Take in account, that
this leads to an increased number of
2
D cycles. Multi-Inject allows to define a
HiRes group to being injected at once, which means the content of the loops is
transferred to the
2
D column before a single
2
D gradient is used for analysis.
1
To inject a HiRes group at once, select Multi-Inject.
The content of the loops is transferred to the
2
D column before a single
2
D
gradient is used for analysis.
Figure 92
Multi Inject for High Resolution sampling