6
Method Parameters
Preview (2D-LC)
2D-LC User Guide
162
Further Graphical Explanation
Further Graphical Explanation of the 2D-LC Preview Window:
The grey triangle illustrates which peaks the peak trigger settings detect in the reference chromatogram. To add
or remove the cut, double-click the grey triangles in the preview.
The grey line in the preview marks the stop time.
In the following example, the
Stoptime
is too short to analyze all cuts. Therefore you must change the stop time to
57 min as indicated by
2
D-gradients. For example use the optimization tool or double click the grey line (green
arrow) marking the current stop time. Then the SW automatically adjusts and takes up the stop time. The alert
icon will disappear (unless anything else is wrong).
NOTE
Hovering over the alert icon gives an idea of what’s wrong.
Marked
cuts
Marked cuts
are displayed either in dark blue bars (can be analyzed) or in dark red bars (cannot be analyzed).
Содержание 1290 Infinity II 2D-LC Solution
Страница 122: ...5 2D LC Data Acquisition in MassHunter Workstation 11 Instrument Details 2D LC User Guide 122 3 Click Screen 4 Click OK...
Страница 263: ...2D LC User Guide 263 9 Data Analysis GC Image Basic Information 3 Import parameters...
Страница 264: ...9 Data Analysis GC Image Basic Information 2D LC User Guide 264 4 Fit view...
Страница 267: ...2D LC User Guide 267 9 Data Analysis GC Image Basic Information 11 View single 2D chromatograms 12 Select blobs...