Agilent 1260 Infinity Quaternary LC VL Manual and Quick Reference
17
Introduction
1
Quaternary pump
Quaternary pump
The quaternary pump is based on a two-channel, dual-plunger in-series design
which comprises all essential functions that a solvent delivery system has to
fulfill. Metering of solvent and delivery to the high-pressure side are
performed by one pump assembly which can generate pressure up to 400 bar.
Degassing of the solvents is done in a built-in vacuum degasser. Solvent
compositions are generated on the low-pressure side by a high-speed
proportioning valve (MCGV).
The pump assembly includes a pump head with a passive inlet valve and an
outlet valve. A damping unit is connected between the two plunger chambers.
A purge valve including a PTFE frit is fitted at the pump outlet for convenient
priming of the pump head.
An active seal wash (optional) is available for applications using concentrated
buffers as solvents.