Agilent 1260 Infinity II Preparative Autosampler User Manual
4
Installation of the Optional Sample Cooler
Install the Sample Cooler
Install the Sample Cooler
Parts required
p/n
Description
Sampler
G7167-60005
Sample cooler
Power cord
5067-6208
Condensate Drainage Kit
G7167-90170
Technical Note - Installation of the Infinity II Cooler Condensate Drainage
Tubing Kit
Preparations
Sampler is installed in the stack.
C A U T I O N
Condensate inside the sample cooler
Damage to the electronics of the module
➔
Before dismounting the sample cooler: For new installation, wait at least 30 min
before switching on the compressor of the sample cooler.
➔
Make sure the power cords are disconnected
➔
Make sure there is no condensate inside the module
N O T E
If the sample cooler is disconnected from the power supply, you should wait for at least
five minutes before replugging and switching on the compressor again.
N O T E
Even under average humidity conditions, a significant amount of condensed water gathers
every day. A suitable container must be provided and emptied regularly in order to avoid
overflow.
N O T E
For best cooling performance of the sampler all drawers must be installed in the module.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...