![Agilent Technologies 1260 Infinity II Скачать руководство пользователя страница 247](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537247.webp)
Agilent 1260 Infinity II Preparative Autosampler User Manual
247
LAN Configuration
13
PC and Agilent ChemStation Setup
Agilent ChemStation Setup
This procedure describes the Agilent ChemStation B.04.02 setup for the
1290 Infinity system using the 1290 Infinity DAD (G4212A) as the interfacing
module. The setup works in the same way for all other systems.
1
Open the ChemStation Configuration Editor.
2
Select from the menu
Configure - Instruments
.
3
Select
Modular 3D LC System
.
4
Give the Instrument a name.
N O T E
The LAN must be connected to detector due to high data load on communication to
Control Software.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...