Agilent 1260 Infinity II Preparative Autosampler User Manual
221
Hardware Information
12
Interfaces
Special Interfaces
There is no special interface for this module.
ERI (Enhanced Remote Interface)
ERI replaces the AGP Remote Interface that is used in the HP
1090/1040/1050/1100 HPLC systems and Agilent 1100/1200/1200 Infinity
HPLC modules. All new InfinityLab LC Series products using the FUSION core
electronics use ERI. This interface is already used in the Agilent Universal
Interface Box 2 (UIB2)
Table 14
ERI signal distribution
Pin
Signal
Description
1
START REQUEST
(L) Request to start injection cycle (for example, by start key on
any module). Receiver is the autosampler.
2
STOP
(L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is
any module performing run-time controlled activities.
3
READY
(H) System is ready for next analysis. Receiver is any sequence
controller.
4
POWER ON
(H) All modules connected to system are switched on. Receiver is
any module relying on operation of others.
5
Not used
6
SHUT DOWN
(L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
7
START
(L) Request to start run / timetable. Receiver is any module
performing run-time controlled activities.
8
PREPARE
(L) Request to prepare for analysis (for example, calibration,
detector lamp on). Receiver is any module performing
pre-analysis activities.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...