Agilent 1260 Infinity II Preparative Autosampler User Manual
167
Maintenance
9
Exchange the Rotor Seal
Exchange the Rotor Seal
When
When poor injection volume reproducibility or when injection valve is leaking.
Tools required
p/n
Description
8710-0510
Wrench open 1/4 — 5/16 inch
8710-2394
Hex key 9/64 inch 15 cm long T-handle
Cleaning tissue and appropriate solvent like isopropanol or methanol
Parts required
p/n
Description
5068-0268
Rotor Seal
5068-0267
Stator
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running.
C A U T I O N
Reduced life time of the injection valve
Component cleanliness is crucial for the life time of the injection valve.
➔
Replace the rotor seal in a clean environment.
C A U T I O N
Removing the stator head
The stator face is held in place by the stator head. When you remove the stator head,
the stator face can fall out of the valve.
➔
Carefully handle the valve to prevent damage to the stator face.
➔
Carefully handle the stator face during sonication.
Содержание 1260 Infinity II
Страница 24: ...24 Agilent 1260 Infinity II Preparative Autosampler User Manual 1 Introduction Leak and Waste Handling ...
Страница 38: ...38 Agilent 1260 Infinity II Preparative Autosampler User Manual 3 Using the Module Turn on off 5 ...
Страница 188: ...188 Agilent 1260 Infinity II Preparative Autosampler User Manual 9 Maintenance Replace the Module Firmware ...
Страница 271: ...Agilent 1260 Infinity II Preparative Autosampler User Manual 271 Index ...