Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Install the Integrated Column Compartment
Install the Integrated Column Compartment
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
#
p/n
Description
1
G7130-60030
ICC Column Heater 3 µL volume
OR
1
G7130-60060
ICC Column Heater 6 µL volume
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Column is installed in the ICC
•
Open the doors of the sampler
WA R N I N G
Fire or explosion due to missing leak drainage
Solvents may run into the module if Column Shelf or Integrated Column
Compartment are not installed.
➔
Install either Column Shelf or Integrated Column Compartment before starting the
module.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...