![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552040.webp)
40
Agilent InfinityLab LC Series Vialsamplers User Manual
2
Site Requirements and Specifications
Physical Specifications
Physical Specifications
Table 1
Physical Specifications
Type
Specification
Comments
Weight
19 kg (41.9 lbs)
w/o Thermostat
Dimensions
(height × width × depth)
320 x 396 x 468 mm
(12.8 x 15.6 x 18.4 inches)
Line voltage
100 – 240 V~, ± 10 %
Wide-ranging
capability
Line frequency
50 or 60 Hz, ± 5 %
Power consumption
350 VA / 350 W / 1195 BTU/h
Ambient operating
temperature
4 - 40 °C (39 - 104 °F), without sample
cooler or sample thermostat up to 55 °C
(131 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
1
1
If a sample cooler or thermostat is included the upper value for humidity can be reduced. Please
check your lab conditions to stay beyond dew point values for non–condensing operation.
Non-condensing
Operating altitude
Up to 3000 m (9842 ft)
Non-operating altitude
Up to 4600 m (15092 ft)
For storing the module
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution degree 2
For indoor use only.
ISM Classification
ISM Group 1 Class B
According to CISPR 11
Permitted solvents
Auto-ignition temperature
≥
200 °C
Boiling point
≥
56 °C
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...