Agilent InfinityLab LC Series Vialsamplers User Manual
191
Maintenance and Repair
9
Remove and Install Doors
Remove and Install Doors
When
If the front door is defective or the hinge are damaged.
Tools required
Description
Flat screwdriver
Parts required
p/n
Description
G7129-68702
Door Assy
Preparations
Finish any pending acquisition job and return any plate on the workspace back to the hotel.
1
Press the release buttons and pull the front door out.
2
For the Installation of the front door, insert the hinges
into their guides and move the door in until the release
buttons click into their final position.