Error Information
8
Integrated Column Compartment (ICC) Heater Error Messages
Integrated Column Compartment (ICC) Heater Error Messages
ICC Temp Wait Timeout Error
Error ID: 30820
The Integrated Column Compartment (ICC) Heater has detected that the left
heater control has reached a time limit.
ICC Heater Overcurrent Failure
Error ID: 30821
The Integrated Column Compartment (ICC) Heater has detected an
overcurrent in the heater circuits. For safety reason heater over-current
protection became active.
Probable cause
Suggested actions
1
Defective cable or connector
Please contact your Agilent service
representative.
2
Defective ICC heater
Replace the ICC heater.
Probable cause
Suggested actions
1
Defective cable or connector
Please contact your Agilent service
representative.
2
Defective ICC heater
Replace the ICC heater.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...