8
Error Information
Sampler Error Messages
Missing Wash Vial
Error ID: 34306
The wash vial programmed in the method was not found.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Invalid Vial Position
Error ID: 34309
The vial position defined in the method or sequence does not exist.
The reflection sensors on the transport assembly flex board are used to
automatically check which sample trays are installed (coding on tray). If the
vial position does not exist in the current sample tray configuration, the error
message is generated.
Probable cause
Suggested actions
1
No wash vial in the position defined in the
method.
Install the wash vial in the correct position, or
edit the method accordingly.
Probable cause
Suggested actions
1
Incorrect tray or trays installed.
Install the correct trays, or edit the method or
sequence accordingly.
2
Incorrect vial positions defined in the
method or sequence.
Check method settings vs. installed tray.
3
Tray recognition defective (dirty sample tray
or defective transport assembly flex board).
Ensure the coding surfaces of the sample tray
are clean (located at the rear of the sample
tray). If this does not solve the problem,
contact your Agilent service representative.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...