Agilent InfinityLab LC Series Vialsamplers User Manual
8
Error Information
Sampler Error Messages
Initialization Failed
Error ID: 34392, 34391
The sampler failed to complete initialization correctly.
The sampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Probable cause
Suggested actions
1
Mechanical obstruction.
Ensure unobstructed movement of the
transport assembly.
2
Defective needle station board.
Please contact your Agilent service
representative.
3
Defective transport assembly flex board.
Please contact your Agilent service
representative.
4
Defective hydraulic unit motor.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...