![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 160](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552160.webp)
160
Agilent InfinityLab LC Series Vialsamplers User Manual
7
Troubleshooting and Diagnostics
Troubleshooting Guide for the Sample Transport Assembly
Poor alignment
With Error messages
• motor overtemp
(0 or 2 or 3)
• movement failed
(0 or 2 or 3)
1
Start automatic referencing in Lab Advisor. For this procedure it is
mandatory to use the calibration tool.
On completion of the procedure, the corrected values are stored in the
instrument firmware.
2
Exchange the gripper arm assembly (Gripper assembly (G1313-60010)), see
3
Exchange Transport assembly ( Sample Transport Assembly
(G7129-60600)).
4
Exchange the Autosampler Main Board ( Autosampler Specific Board
(G7129-65800)) or the fusion board ( Fusion Board (G4200-65880)).
5
For all three steps (step 2, 3, and 4) you have to do an automatic referencing
again to store the correct values in the instrument firmware again.
WA R N I N G
Personal damage, damage to the module
➔
Some of these procedures require a trained service engineer. Persons who are not
qualified MUST NOT perform these procedures.
N O T E
When a motor over temperature message has occurred, the sampler must be turned OFF
for about 10 min to allow the motor to cool down.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...