![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 158](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552158.webp)
158
Agilent InfinityLab LC Series Vialsamplers User Manual
7
Troubleshooting and Diagnostics
Troubleshooting Guide for the Sample Transport Assembly
Troubleshooting Guide for the Sample Transport Assembly
This troubleshooting guide is meant to help you diagnose and repair
autosampler problems.
In general, autosampler problems can be divided into three categories.
1
Intermittent lock-ups with or without vial in the gripper fingers with error
messages
Many times the sampler is being used very heavily.
•
motor overtemp
(0 or 1 or 2 or 3)
•
movement failed
(0 or 1 or 2 or 3)
•
missing vial
2
Jittery (shaky) movement in X and/or theta axes and/or when the needle
goes through the gripper arm into the vial with error messages
•
motor overtemp
(0 or 2)
•
movement failed
(0 or 2)
3
Poor alignment, seen during vial pickup and vial replacement and/or when
the needle hits the gripper arm with error messages
•
motor overtemp
(0 or 2 or 3)
•
movement failed
(0 or 2 or 3)
•
missing vial
N O T E
Motor 0=X; 1=Z; 2=Theta; 3=Gripper.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...