![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 155](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552155.webp)
Agilent InfinityLab LC Series Vialsamplers User Manual
155
Troubleshooting and Diagnostics
7
Tests
Tests
ICC Heater Test
The Oven Test is used to evaluate the heating performance of the heating
element. The heating rate over a 10 K interval from the start temperature is
determined. The start temperature must be between 30 °C and 50 °C, and is
determined as follows:
•
If the current oven temperature is below 30 °C, the oven tries to attain a
temperature of 30 °C. 30 °C is used as the start temperature.
•
If the current oven temperature is above 30 °C but below 50 °C, the current
oven temperature is used as the start temperature.
•
If the current oven temperature is above 50 °C, an error message is
displayed. The oven must then be allowed to cool to below 50 °C before the
test can be run.
Sample Cooler Function Test
The Sample Cooler Function Test can be used as a simple verification that the
sample cooler is functioning.
After the test has been started, it begins to acquire data from the cooler's
PT1000 temperature sensor. As soon as the temperature has equilibrated (that
is, the temperature does not change by more than 0.5 °Cover a 10 s period) the
cooler is turned on and measurement is started.
For the test to succeed, three temperature checkpoints must be reached within
a specified time.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...