Thermal Fuse Open
Error ID: 2680
The thermal fuse of the optical unit heater has failed.
Heater Resistance Too High
Error ID: 2681
The resistance of the heater foil is above the set limit.
Probable cause
Suggested actions
1
Heater cable disconnected.
Please contact your Agilent service representa-
tive.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective thermal fuse.
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Heater cable disconnected.
Please contact your Agilent service representa-
tive.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective heater.
Please contact your Agilent service representa-
tive.