Error Information
General Error Messages
Leak Sensor Short
Error ID: 0082
The leak sensor in the module has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak sensor current to
change within defined limits. If the current increases above the upper limit, the
error message is generated.
Compensation Sensor Open
Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor increases above the upper limit, the
error message is generated.
Probable cause
Suggested actions
1
Defective leak sensor.
Please contact your Agilent service representa-
tive.
2
Leak sensor incorrectly routed, being
pinched by a metal component.
Please contact your Agilent service representa-
tive.
3
Power switch assembly defective
Please contact your Agilent service representa-
tive.
4
Cable or contact problem.
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Loose connection between the power switch
board and the mainboard
Please contact your Agilent service representa-
tive.
2
Defective power switch assembly
Please contact your Agilent service representa-
tive.