Agilent InfinityLab LC Series RID User Manual
100
7
Error Information
Refractive Index Detector Specific Error Messages
Lamp Current too Low
Error ID: 2694
Wait Function Timed Out
Error ID: 2692
Wait for temperature or wait for defined signal has not been fulfilled within the
specified time frame.
Probable cause
Suggested actions
1
Optical unit cable disconnected.
Please contact your Agilent service representa-
tive.
2
Defective main board.
Please contact your Agilent service representa-
tive.
3
Defective lamp or optics.
Please contact your Agilent service representa-
tive.
Probable cause
Suggested actions
1
Time too short.
Increase time.