Agilent 1260 Infinity DAD and MWD User Manual
133
Error Information
7
General Error Messages
2
If it is desired that the system has only one leak drain tube, then it's
possible in increase the length of the small plastic tube which is connected
to port# 4 of the sampler switching valve. This tube can then be taken to
waste separately. The tube which normally serves as the detector cell outlet
tube can be used for this purpose.
Open Cover
Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Cover Violation
Error ID: 7461
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed while the lamps are on (or if an attempt is made to switch on
for example the lamps with the foam removed), the lamps are switched off,
and the error message is generated.
Probable cause
Suggested actions
1
The top foam was removed during
operation.
Please contact your Agilent service representative.
2
Foam not activating the sensor.
Please contact your Agilent service representative.
3
Defective sensor or main board.
Please contact your Agilent service representative.
Probable cause
Suggested actions
1
The top foam was removed during
operation.
Please contact your Agilent service representative.
2
Foam not activating the sensor.
Please contact your Agilent service representative.
Содержание 1260 Infinity G1315C
Страница 1: ...Agilent Technologies Agilent 1260 Infinity Diode Array and Multiple Wavelength Detector User Manual ...
Страница 18: ...18 Agilent 1260 Infinity DAD and MWD User Manual 1 Introduction Bio inert Materials ...
Страница 98: ...98 Agilent 1260 Infinity DAD and MWD User Manual 4 Using the Detector Solvent Information ...
Страница 162: ...162 Agilent 1260 Infinity DAD and MWD User Manual 8 Test Functions Wavelength Verification and Calibration ...
Страница 221: ...Agilent 1260 Infinity DAD and MWD User Manual 221 Parts for Maintenance 10 Nano Flow Cells Figure 76 Content of kits ...
Страница 228: ...228 Agilent 1260 Infinity DAD and MWD User Manual 10 Parts for Maintenance Accessory Kits ...
Страница 260: ...260 Agilent 1260 Infinity DAD and MWD User Manual 12 Hardware Information Early Maintenance Feedback EMF ...
Страница 290: ...290 Agilent 1260 Infinity DAD and MWD User Manual 13 LAN Configuration Manual Configuration ...
Страница 313: ......