Calibrating Probes
4
N2887A & N2888A Calibration and Deskew Procedure
1168/9A Probes User’s Guide
133
N2887A & N2888A Calibration and Deskew Procedure
The N2887- 60002 Calibration/Deskew kit contains the
following parts. The last three items in the list are required
for the calibration procedure.
■
BNC (male) to SMA (male) adapter
■
SMA (male) to BNC (female) adapter
■
SMA (male) to SMA (male) adapter
■
50
Ω
SMA terminator
■
Deskew Fixture
Calibration of the N2887A/N2888A probe heads consists of a
vertical calibration and a skew calibration. The vertical
calibration should be performed before the skew calibration.
Both calibrations should be performed for best probe
measurement performance.
NOTE
Before calibrating the probe, verify that the Infiniium oscilloscope has
been recently calibrated and that the calibration change in temperature is
±
5
o
C. If this is not the case, calibrate the oscilloscope before calibrating
the probe. This information is found in the Infiniium Calibration dialog box.
Содержание 1168/9A
Страница 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Страница 5: ...Contents 1168 9A Probes User s Guide 5 N5381A and N5382A Probe Heads 248 Other Accessories 249 ...
Страница 6: ...6 1168 9A Probes User s Guide Contents ...
Страница 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Страница 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Страница 69: ...Using Probe Heads 2 Recommended Configurations 1168 9A Probes User s Guide 69 Figure 36 Removing the Probe ...
Страница 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Страница 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Страница 136: ...136 1168 9A Probes User s Guide 4 Calibrating Probes N2887A N2888A Calibration and Deskew Procedure ...
Страница 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Страница 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Страница 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...