Calibration Test Record
The AC Accuracy range is based on the Infiniium oscilloscope’s accuracy.
Calculate the minimum and the maximum limits corresponding to the
accuracy of your Infiniium oscilloscope.
Agilent Technologies
1147A DC to 50 MHz Current Probe
Recommended Test Interval: 1 year
Serial No.: ___________________
Certification Date: ______________
Tested By: ___________________
Certification Temperature: _______________________
Recommended Date of Next Certification: ___________
Test
Limit
Minimum
Results
Limit
Maximum
(1) AC Accuracy [Arms]
9.75
10.25
NOTE