1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Pressure Above Upper Limit
The system pressure has exceeded the upper pressure limit.
Probable Causes
•
Upper pressure limit set too low.
•
Blockage in the flowpath (after the damper).
•
Defective damper.
•
Defective CSM board.
Suggested Actions
✔
Ensure the upper pressure limit is set to a value suitable for the analysis.
✔
Check for blockage in the flowpath.
✔
Exchange the damper.
✔
Exchange the CSM board.