1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Restart Without Cover
The nano pump was restarted with the top cover and foam open.
The sensor on the CSM board detects when the top foam is in place. If the
nano pump is restarted with the foam removed, the nano pump switches off
within 30 s, and the error message is generated.
Probable Causes
•
The nano pump started with the top cover and foam removed.
•
Rear of the module is exposed to strong direct sunlight.
Suggested Actions
✔
Replace the top cover and foam.
✔
Ensure that the rear of the module is not exposed to strong sunlight.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......