1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Leak Sensor Open
The leak sensor in the nano pump has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
•
Leak sensor not connected to the CSM board.
•
Defective leak sensor.
•
Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions
✔
Ensure the leak sensor is connected correctly.
✔
Exchange the leak sensor.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Страница 4: ...iv 1100 Series Nano Pump Service Manual ...
Страница 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Страница 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Страница 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Страница 174: ...162 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 27 Nano pump hydraulic path 1 9 3 6 5 8 7 10 4 2 ...
Страница 176: ...164 1100 Series Nano Pump Service Manual 4 Parts and Materials Figure 28 Pump Head Assembly 1 2 3 5 6 7 11 9 4 8 10 ...
Страница 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Страница 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Страница 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Страница 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Страница 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Страница 287: ......