Troubleshooting and Test Functions
4
1100 Series FD Reference Manual
99
Flow Cell Removed
The detector has an automatic cell recognition system. When the flow cell is
removed, the lamp is turned off and a NOT READY condition exists. If the flow
cell is removed during an analysis, a SHUT DOWN is generated.
Probable Causes
•
Flow cell has been removed during analysis.
Suggested Actions
✔
Insert flow cell and turn on the lamp.