1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
Flash Trigger Lost
This message is displayed when the flash trigger is no longer generated.
Probable Causes
•
firmware problem
•
defective encoder
Suggested Actions
✔
Reboot the detector (power cycle).
✔
Multi Mode Off: exchange the firmware or replace the main board.
✔
Multi Mode Ex: replace EX monochromator assembly.
✔
Multi Mode Em: replace EM monochromator assembly.