200
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Replacing the Detector’s Firmware
The installation of new firmware is required:
•
if new version solves problems of currently installed version, or
•
if after exchange of the main board (FLM) the version on board is older
than the previously installed one.
To upgrade the detector’s firmware the following steps have to be taken:
1
Load the firmware into the detector, see help system of your user interface.
2
If FLM board was replaced, reenter the serial number information of the
module through the user interface, see
“Entering the Serial Number using
on page 161 or
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......