Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
187
Replacing the EM and EX Grating
When required
If defective
Tools required
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
Parts required
EM Grating assembly G1321-60004 (
includes
protection ring)
EX Grating assembly G1321-60003 (
includes
protection ring)
C A U T I O N
When working on the optical unit, a clean workbench with ESD protection mat must be
available. Otherwise optical components or electronic boards may be damaged.
N O T E
The grating assembly is shipped with a special transport packaging and includes a
protection ring. DO NOT touch the grating surface and DO NOT bend the protection ring.
Preparations for this procedure:
•
Optical unit has been removed as described in
•
The cables from the position encoders (at the rear of
the EM or EX grating assembly) should be
disconnected.
•
Disconnect the encoder cable of the
EX-monochromator from FLF board.
1
Place the optical on the ESD protection mat and remove
the front foam (A). Locate the EM and EX grating
assembly.
A
B
A
B
A
EM
EX
A
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......