Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
183
Replacing FLL board and Trigger Pack
When required
If defective
Tools required
Screwdriver POZI 1 PT3
Hexagonal key, 4 mm and 2.5 mm
Parts required
FLL Board G1321-66512 (
includes
trigger pack)
C A U T I O N
When working on the optical unit, a clean workbench with ESD protection mat must be
available. Otherwise optical components or electronic boards may be damaged.
WA R N I N G
The instrument should not be operated without Xenon flash lamp. Otherwise the
Triggerpack electronics will be damaged.
Preparations for this procedure:
•
Optical unit has been removed as described in
•
The foam must be in place. Otherwise cutoff filter and
diffuser will fall out.
1
Place the optical unit on the ESD protection mat.
B