160
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
7
Remove the processor board. Place the board on the ESD kit.
8
On the new board check the switch setting of address switch S1, see
“Setting the 8-bit Configuration Switch"
9
Install the new processor board and reconnect the connectors. Assure that
the board is fitted correctly into the board recess in the rear panel.
10
Refit the screws at the REMOTE and GPIB connectors and the nuts of the
ANALOG connectors.
N O T E
When removing connectors, counter-hold with one hand on connector J12.
Figure 59
Location of Connectors on FLM Board
J1 - GPIB
J2 - RS-232C
J3 - REMOTE
J4/5 - Analog output
J6/7 - CAN
J11 - Power supply
J12 - EM Monochromator
J14 - EX Monochromator
J17 - NOT USED
J18 - FLF board
J2
J1
J2
J3
J4
J5
J1
B
S1
J1
J1
J1
J1
J2
J2
J6
J7
N O T E
An incorrect switch setting (for example, TEST/BOOT) may cause the module to revert to a
basic mode (yellow or red flashing status light). In this case turn off the module, reset the
address switches, and turn on the module again.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......