
Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
141
Replacing Leak Handling System Parts
1
Remove the front cover.
2
Pull the leak funnel out of the leak funnel holder.
3
Pull out the leak funnel with the tubing.
4
Insert the leak funnel with the tubing in its position.
5
Insert the leak funnel into the leak funnel holder.
6
Replace the front cover.
When required
If the parts are corroded or broken
Tools required
None
Parts required
Leak funnel 5061-3356
Leak funnel holder 5041-8389
Leak tubing (120 mm) 0890-1711
Figure 57
Replacing Leak Handling System Parts
Leak tubing
Leak funnel
Leak funnel holder
Leak sensor
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......