1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Correcting Leaks
1
Remove the front cover.
2
Use tissue to dry the leak sensor area and the leak pan.
3
Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4
Replace the front cover.
When required
If a leakage has occurred in the flow cell area or at the capillary connections
Tools required
Tissue
Two 1/4 inch wrenches for capillary connections
Parts required
None
Figure 56
Observing for Leaks
Leak sensor