Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
139
Flow Cell Flushing
In case the cell is contaminated, follow the procedure below.
Flushing Procedure
1
Flush with bidistilled water.
2
Flush with nitric acid (65%) using a glass syringe.
3
Leave this solution in the cell for about one hour.
4
Flush with bidistilled water.
When required
If flow cell is contaminated
Tools required
Glass syringe, adapter
Parts required
Bidistilled water, nitric acid (65%), tubings to waste
N O T E
Aqueous solvents in the flow cell can built up algae. Algae do fluoresce. Therefore do not
leave aqueous solvents in the flow cell for longer periods. Add a small percentage of
organic solvents (e.g. Acetonitrile or Methanol ~5%).
C A U T I O N
This concentration of nitric acid is dangerous and proper attention to safety should be
given. Also the nitric acid flushing procedure is not an infallible remedy for a dirty cell.
It is to be used as a last attempt to salvage the cell before cell replacement. Note that
the cell is a consumable item.
N O T E
Do not exceed the pressure limit of 20 bar (0.2 MPa).
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......